Continuous cardiac monitoring in wearable devices demands classifiers that are simultaneously accurate, energy-efficient, and deployable on resource-constrained hardware. While deep neural network approaches have demonstrated high classification accuracy for electrocardiogram (ECG) arrhythmia detection, their substantial parameter counts and reliance on multiply-accumulate-intensive operations make them impractical for low-cost edge platforms. In this work, we propose ECG-LDC, a hardware-software co-design framework that adapts Low-Dimensional Computing (LDC) for real-time ECG arrhythmia classification. ECG-LDC employs a dual-encoder architecture with dedicated value and feature codebooks to independently encode morphological waveform features and RR-interval temporal features, enabling effective capture of both intra-beat and inter-beat cardiac dynamics. The framework encompasses data preprocessing, model training, and a hardware accelerator architecture prototyped on the Pynq-Z2 platform. Implemented using binary representations and XOR/XNOR-based operations, ECG-LDC achieves $97.18\%$ accuracy with a memory footprint of only $3.86\ \text{kB}$. ECG-LDC sacrifices approximately $1.8\%$ accuracy versus SOTA TinyML classifiers but achieves $11$~$ 570\times$ reduction in memory usage; among FPGA-based five-class arrhythmia classifiers, it delivers the highest accuracy with up to $2.4\times$ fewer LUTs and zero DSP block utilization, affirming its suitability for real-time arrhythmia detection on resource-constrained wearable platforms.
Our work presents a method for ECG segmentation and arrhythmia detection using Tiny Machine Learning (TinyML) models for real-time, on-device inference on resource-constrained embedded systems. We develop INT8 quantized autoencoder-based TinyML models with minimal layers and parameters for embedded deployment. These models are evaluated using a custom dataset derived from the MIT-BIH Arrhythmia Database and validated in both PC-based simulations and on-device environments. For the evaluations, over 95,000 ECG segments are processed on an ESP32-S3 microcontroller running the TensorFlow Lite Micro runtime. Post-evaluation, detailed analysis, including annotation-wise and record-wise failure analysis, is conducted to characterize model behavior across diverse ECG morphologies and rhythm patterns and to explain missed detections. In several cases, apparent misclassifications may correspond to early or subtle anomaly patterns labeled as normal in the reference annotations, highlighting the model's sensitivity. A refined evaluation by filtering out ambiguous cases in the dataset shows that the best-performing DNN-based autoencoder achieves a recall of 84%, an F1-score of 79%, a model size of approximately 180 KB, and an inference latency of 9 ms on-device. These results demonstrate the feasibility of low-power, privacy-preserving embedded wearable systems capable of performing accurate arrhythmia detection entirely on-device.